Key Features
- Superior accuracy within ±0.25% of reading ±1°C
- Large 51 mm (2”) aperture diameter
- High effective emissivity 1.0 at 0.7 to 1.8um, 0.998 typical at other wavelengths
- Rugged housing designed for fixed installation in laboratory or test department
- Manufactured and tested to meet rigid quality control standards
- Furnished with certificate of calibration traceable to NIST
- RS232 (standard) or RS485 (optional) serial communications port
Details
A high standard, precision blackbody calibration source designed with a unique uniformly heated spherical cavity that achieves a near ideal emissivity of 0.998 or better.
The M300 will deliver any temperature between 200 °C and 1150 °C (392° TO 2102 °F) – accurately. A convenient microprocessor-based, self-tuning digital PID controller holds the selected temperature to within 1 °C, thus assuring high accuracy calibration.
Typical Calibration Applications
- Infrared Temperature Sensors
- Infrared Thermal Imaging Systems
- Spectrographic Analyzers
- Spectral Radiometers
- Heat Flux Meters