- Relative Humidity Range: 10 to 95%RH
- Relative Humidity Resolution: 0.01 %RH
- Relative Humidity Uncertainty: (@ PcTc *) 0.5 %RH
- Chamber Temperature Range: 10 to 60 ºC
- Chamber Temperature Resolution: 0.01 ºC
- Chamber Temperature Uniformity: ** ±0.1 ºC
- Chamber Temperature Uncertainty: * 0.05 ºC
- Chamber Pressure Range: Ambient
- Gas Flow Rate Range: 5 to 10 l/m
*Represents an expanded uncertainty using a coverage factor, k=2, at an approximate level of confidence of 95%.
** When operated at temperatures within 10 ºC of room ambient temperature.
Features
- 0.5 %RH Uncertainty 1
- Traceable to SI 3
- Based on NIST Proven Two-Pressure Principle
- Generate: RH, DP, FP, PPM, Multipoint Profiles
- Computerized Internal Transducer Calibration
- Computes System Uncertainties in Real Time
- Automatically Applies Enhancement Factors
- No Refrigerants - Thermoelectric Cooling/Heating
- Only 4 square feet of floor space (20" x 30 ")